How to Read a Mass Spec (LC-MS) Identity Result on a Peptide COA
By Peptides & Coffee · Updated Nov 18, 2026 · About 6 minutes
An identity result answers one question: is the main material the molecule it is labeled as? Mass spectrometry (often LC-MS or ESI-MS) measures the mass of the molecules in the sample. The certificate compares the observed mass with the expected (theoretical) mass of the declared peptide. A close match supports identity; it says nothing about how much peptide is in the vial or how pure it is, which are separate results.
What an identity test is for
HPLC purity tells you how much of the UV-detected material sits in the main peak. It can't tell you what that peak is. A sample could be highly pure and still be the wrong molecule. The identity result fills that gap, which is why a complete certificate reports identity next to purity and measured amount. The basics of each section are in how to read a peptide COA.
How mass spectrometry measures identity
In LC-MS the sample first passes through a liquid chromatography column, then into a mass spectrometer. Electrospray ionization (ESI) gives the peptide molecules an electric charge, and the instrument measures their mass-to-charge ratio (m/z). From those readings the lab works out the molecular mass of the main component and compares it with the mass calculated from the declared sequence.
Some labs use MALDI-TOF instead; the idea is the same, comparing a measured mass with the expected one.
Reading expected vs. observed mass
| Expected (theoretical) mass | The mass calculated from the declared peptide's formula, usually in daltons (Da). Labs may state the monoisotopic or the average mass; the two differ slightly for the same molecule. |
|---|---|
| Observed mass | The mass the instrument found for the main component. |
| Difference | Sometimes given in Da or in parts per million (ppm). Small differences are normal and depend on the instrument; the lab's stated acceptance window is what counts. |
| Result | Often written "Conforms", "Consistent with structure", "Pass" or "Identity confirmed". |
To read it: check that the expected mass belongs to the compound named on the certificate (public databases such as PubChem list molecular weights), check that the observed mass is close to it, and check that the result is for your lot.
Charge states and multiple peaks
Larger peptides often pick up more than one charge, so the spectrum can show several peaks, written for example as [M+2H]2+ or [M+3H]3+. Each is the same molecule seen at a different m/z. The lab combines them into one molecular mass. Extra peaks that don't fit the main molecule can be impurities, adducts such as sodium, or fragments; the certificate should say which result it relies on.
How identity is written on certificates
- Full detail: expected mass, observed mass and a spectrum image. The most informative form.
- Pass/fail only: "Identity: Conforms (MS)". Acceptable, but you can't check the numbers yourself.
- No method named: "Identity: Pass" with no method. Ask which method was used.
- Measured against the declared compound: some labs quantify the declared peptide against a reference standard and report identity that way rather than by a separate MS line.
Our lot pages show the identity result exactly as the certificate states it, or "Not reported" when it doesn't state one. We don't fill in missing results.
What identity doesn't prove
- It doesn't measure purity or the amount per vial; see purity and net content.
- Some closely related forms have very similar or identical masses, so mass alone can't separate every variant.
- It describes the sample tested, so it only applies to your vial if the lot matches; see how to check a lot number.
- It only counts if the lab's own record shows it. Paste the certificate's code into the free COA check to open that record.
Check a COA now
Paste a lab verify link or code. Free, nothing is stored.
FAQ
What does identity mean on a peptide COA?
It is the result of a test that checks whether the main material is the molecule named on the certificate. It is usually done by mass spectrometry, comparing the observed mass with the expected mass.
What is LC-MS on a COA?
Liquid chromatography–mass spectrometry. The sample is separated on a column and the mass spectrometer measures the mass-to-charge ratio of the molecules, from which the lab works out the molecular mass of the main component.
How close should observed and expected mass be?
Close enough to fall inside the lab's stated acceptance window, which depends on the instrument. Small differences in Da or ppm are normal.
Why are there several peaks on a mass spectrum?
Peptides can carry more than one charge, so the same molecule appears at several m/z values. Other peaks can be adducts, fragments or impurities.
Does an identity pass mean the peptide is pure?
No. Identity confirms what the main material is. Purity (HPLC) and the measured amount per vial are separate results.
What if my COA has no identity result?
Then the certificate doesn't confirm what the material is. Our lot pages show "Not reported" in that case rather than filling in a result.
Related guides
This guide explains how to read and check analytical chemistry documents. It is not advice on using any product. Products referenced on this site are for laboratory research use only and are not for human or veterinary consumption.